Three-dimensional roll bag

Core Uses

Medical device sterilization packaging, achieving sterilization and bacteria barrier.
Aseptic storage and transportation to prevent secondary contamination.
Suitable for irregular/small devices (e.g., dental handpieces, scissors, catheters, etc.), preventing extrusion and deformation.
Direct aseptic clinical access with transparent and easy-tear design, improving efficiency.

Standard Usage

Preparation: Select compliant roll pouches; clean and dry devices, and cover sharp ends with protective sleeves.
Cutting: Cut according to device size + 2–3 cm sealing edge, retaining the three-dimensional fold.
Loading: Place devices in the center with sharp ends facing the pouch bottom; do not exceed 2/3 of the pouch volume.
Heat sealing: Seal with a medical heat sealer.
Sterilization: Sterilize by the corresponding method and confirm the indicator card changes color to qualified.
Storage: Store in a dry and ventilated place, follow the first-in-first-out principle, and avoid extrusion and damage.